I.
EQUIPMENT SETUP
For this lab, we will be using the PowerLab Digital Voltmeter (DVM
- accessible through Chart) to monitor the potential between the
two compartments of the Ussing chamber. In the
"short-circuit" configuration we will add a second "circuit" to
the Ussing chamber to provide a countercurrent flow. This
circuit will consist of a DC source (2 C batteries), a set of
potentiometers to regulate the electrical current flow, and a
microammeter to measure this imposed current flow.
A small digression on electrodes is in order here.
Assuming we have two wire leads coming off of a recording or
stimulating circuit, what is an appropriate way to make contact
with the fluid on the two sides of the membrane? On first
thought, it might seem sufficient to just immerse the free end
of each wire in the fluid. If this is done, a potential
difference will be observed between the wires, but it will be
erratic and bear little relationship to the actual potential
difference between the two compartments. When a bare metal
electrode is immersed in a saline solution, an electrical
potential is established between the metal and the solution, due
to the metal yielding ions to the solution, or the solution
plating ions onto the metal. This is called electrode or
junction potential, constitutes a variable battery, and is the bane of electrophysiology.
If the junction potentials at the two electrodes were constant
and identical, they would cancel out in the overall potential
measurement. In order for this to happen, both the
conditions at the surface of each electrode, and the ionic
makeup of the solution surrounding the electrode would have to
be constant and identical. However, as current flows in a
single direction through each junction, the properties of the
junction change, e.g. as a result of electroplating of ions onto
the electrode. This process is compounded when electrodes
are used to pass externally applied direct currents. Under
such conditions, bare metal electrodes can rapidly "polarize"
and block further current flow. To minimize these effects,
we will be using metallic silver electrodes which have been
electrically coated with a layer of silver chloride. When
placed into a NaCl solution, for example, the Ag/AgCl/NaCl
interface remains fairly stable and can accurately record and
pass direct (unchanging) currents. |
The
equipment needed for this experiment includes an Ussing chamber,
Ag/AgCl electrodes, a microammeter, a potentiometer box, two dry
cell batteries, and a PowerLab/PC recording station.
Equipment Setup Procedure:
1) Rinse
out the two compartments of the Ussing chamber with tap water,
followed by distilled water, followed by a small amount of frog
Ringer's. Avoid touching the inner surface of the Ussing
chamber blocks. Clean any crusted salt off of the base clamp
and the spacer blocks.
2)
Obtain four chlorided silver post electrodes from the
instructor. Keep them immersed in the frog Ringer's solution in
a small beaker until you are ready to record.
3)
Launch Chart. Turn off all channels except Channel 1.
Set Channel 1 for bipolar recording (activate both the
positive and negative input leads) at a sampling rate of about 4
per second and with an amplifier display range of 200mV.
4)
Select DVM:Channel 1 from the Windows menu. Enlarge the
DVM window by pulling down on the lower right corner.
The DVM will simply reflect the voltage present across
the + Channel 1 input leads whenever Chart is recording.
5)
Assemble the Ussing chamber without any membrane. Fill the
chamber with 9.0cc of frog Ringer's. Place a Ag/AgCl electrode
in the innermost holes on each side of the chamber. Connect
these electrodes to the "+" and "-" inputs of CH1 on the
PowerLab, connecting "+" to side A and "-" to side B
of the chamber.
6)
Record the DVM electrical potential in mV between the two
electrodes. This is a potential is produced by differences
between the two electrodes and will have to be subtracted
from all subsequent voltage measurements to isolate potentials
generated by the frog skin itself. This "electrode
potential" is also the "zero" or baseline voltage at which
you will be holding the preparation for Section III parts B-E.
You may want to try various pairs of your four electrodes and
select the pair which produces the smallest baseline electrode
voltage.
7) Label
the dry ends of your two electrodes A and B with
small pieces of tape, unclip them, then return
them to their original Ringer's container. If you fail to
unclip them at this point, the "mystery metal" of the clips will
rapidly electroplate onto the electrodes and ruin them.
Drain and rinse the Ussing chamber.
II.
DISSECTION AND MOUNTING OF THE FROG SKIN
For this and future experiments involving animal surgery, always
wear surgical gloves, change your gloves whenever they become
torn or contaminated. As soon as you are done handing the frog
skin, remove your gloves and wash your hands thoroughly . |
Procedure:
1)
Obtain a decapitated or pithed frog from the instructor.
2) Use
your scissors to remove a piece of skin from the ventral surface
of the frog at least 2.5cm in diameter.
3)
Return the remains of your frog to an ice chest when you are
through, or give it to another group of students.
4)
Carefully mount the frog skin between the blocks off the Ussing
chamber, using just enough clamp pressure to hold it in place.
Make sure that there are no holes or gaps in the skin
which could allow direct flow of fluids between the two
compartments. The skin should be mounted so that the exterior
(mucosal, apical, green) surface faces the B compartment and the
interior (serosal, basal, white) surface faces the A
compartment. If you get the polarity reversed, do NOT
remount the frog skin, just relabel the compartments.
5) Fill
one chamber with 4.5cc of frog Ringer's and check for leaks.
Then fill the other chamber with 4.5cc of frog Ringer's. Slowly
bubble 10cc of air into each compartment of the Ussing chamber,
using the 10cc syringe and the holes farthest from the
epithelium.
III. RECORDING FROM THE FROG SKIN
The voltage recorded across the frog skin (transepithelial
potential) will reflect not only the active transport of cations
(predominantly Na+), but also the passive flow of
anions (predominantly Cl-) across the epithelium.
Since Cl- ion concentrations are equal on the two
sides of the membrane, and assuming that Cl- ions are
not actively pumped, any Cl- fluxes must be driven by
the electrical potential across the membrane. If we could eliminate this transepithelial potential, by applying an electrical
countercurrent, Cl- ions would cease to flow.
Furthermore, the electrical countercurrent which we apply across
the membrane to hold it at a zero potential, would exactly match
(with opposite polarity) the net current due to active transport
of cations. This follows from Kirchoff's Current Laws -
essentially stating that the same current must flow at all
points in a simple, unbranched circuit.
In order to "short-circuit" the epithelium, pin it to zero
potential, and measure the current required to do so we must
construct a second circuit across the epithelium, containing a
steady voltage source, a potentiometer (variable resistor) to
regulate the current in this circuit, and a microammeter to
measure the current. According to Ohm's law, for resistive
current:
V = IR
where V is the potential in volts, I is the
current in amps, and R is the resistance in ohms.
With a fixed voltage source V (in our case a battery), the current
I flowing through the circuit can be varied by changing the
resistance R at the potentiometer. We will use this
short-circuit countercurrent to quantify sodium transport for
the bulk of this laboratory exercise. |
As you
work through the following procedures, keep three things in
mind:
1)
Fluid levels in the two chambers must be exactly equal, so
that there is no hydrostatic pressure between the two chambers
whenever you are recording an electrical potential.
2) At
least once every 5 minutes during the following experiments you
must slowly bubble 10cc of air into each compartment of the
Ussing chamber, to keep the epithelium oxygenated and allow
active transport processes to continue.
3) This
preparation has a limited viability. Work as rapidly as you
can. Answer short questions as you go, and make sure that you
have sufficient data for the required calculations. However,
wait to perform the longer calculations until after the
experiment is complete.
A.
Measuring the Electrical Potential Generated by Active Transport
1)
Return your two recording electrodes to the chamber and connect
them to the PowerLab box, as before. Be careful that the
electrodes never touch the frog skin.
2)
Activate
the DVM, by clicking on the Start button.
Q1: Which side of the skin is positive
relative to the other? What does this tell you about the likely
direction of sodium pumping when there are no ionic gradients
between the two chambers.
3)
Inactivate
the DVM when you are through.
4)
Remember to periodically bubble air through the two
compartments.
B.
Measuring the Active Transport Current
1)
Insert the second pair of electrodes in the central (smallest
bore) holes of each compartment of the Ussing chamber. Make
sure that these electrodes do not contact the other electrode
pair. Using the wires with alligator clips assemble a
circuit running from one of these electrodes, through the
battery, through the potentiometer box, through the microammeter,
and to the other electrode.
2)
Adjust the
potentiometers to produce the lowest possible current flow, as
measured by the microammeter.
3)
Again, remember to periodically bubble air through the two
compartments.
4)
Now activate the DVM. Adjust the potentiometers to
increase the countercurrent until the voltage across the frog
skin is zero. Remember that you must subtract the DC electrode
voltage from the DVM reading to obtain the true voltage
across the skin. If increasing the applied current increases
the potential, reverse the connections to your two
current-supplying electrodes.
5)
Record the
current required to produce a zero transepithelial potential
(again, remember to take into account your measured baseline
electrode potential).
6)
Inactivate
the DVM when you are through, and drop the applied
countercurrent to its lowest value.
Q2: Does the shortcircuit
countercurrent flow in the same or opposite direction as the
active transport current? Explain why.
Assume
that this short circuit current is required to supply one
electron for every net monovalent cation pumped across the
membrane. Assume further that Na+ is
the only cation effectively displaced by the pump. The net
transport rate of Na+ across
the epithelium may be calculated from the following definitions:
1
microampere = 10-6 amperes
1
ampere = 1 coulomb/second
96,500 coulombs = 1 mole of monovalent charge
carriers
(either electrons,
Na+ ions,
or K+ ions)
It follows
that each µamp of short circuit current is equivalent to 1.04 x
10-11
moles
per second of net cation transport across the membrane. Do the
calculations for the following questions after you have
finished with the experiment:
Q3: Given your recorded short-circuit
current, calculate the net Na+flux
(in moles/sec) across the epithelium in your Ussing chamber.
Q4: The diameter of the main Ussing chamber is
12.5mm. Calculate the net Na+ flux
per unit surface area (moles/sec mm2)
due to active transport.
Q5: Given a volume of 4.5ml in each compartment
of the Ussing chamber, how long would it take to change the
concentration of Na+ on
either side of the membrane by 1mM? Does your result justify
our initial assumption that ionic concentrations in the two
chambers could be treated as invariant?
C.
Pumping Against Ionic Gradients and Cation Specificity of the
Pump
In the
experiments so far, the concentration of ions on each side of
the frog skin has been identical. However, in the living frog,
sodium is pumped against a substantial gradient. We will verify
that sodium is carrying the active transport current and study
the concentration dependence by substituting K+ for
Na+ on the external side of the epithelium, while
keeping the transepithelial potential short-circuited to zero.
1)
Activate the DVM and record the short-circuit current
necessary to zero the transepithelial potential.
2)
Now, using
3cc syringes, withdraw 2.25 cc of Ringer's from the outside
compartment of the Ussing chamber and immediately replace it
with 2.25cc of K-substituted Ringer's (a frog Ringer's solution
in which NaCl has been replaced with KCl). Bubble air through
the compartment to oxygenate and mix the solution. This
solution will now contain 1/2 of the initial Na+
concentration.
3)
Record the
short circuit current under this condition.
4)
Repeat
steps 2&3 to produce outside Na+ concentrations of
1/4, 1/8, 1/16, and 1/32 of the original concentration,
recording the required short-circuit current under each
condition.
5)
When you have made the final measurement, inactivate the DVM
and set the countercurrent to the lowest value. Remove both
pairs of electrodes and gently flush both compartments of
the chamber out with standard frog Ringer's. Do not tip the
chamber over, simply extract the fluid in each side with a
syringe and gently replace it with fresh frog Ringer's.
Repeat the flushing process at least twice.
6)
Immediately refill
both sides with 4.5ml of standard frog Ringer's and return the
electrodes to the chamber. |