I. SETUP
A.
Snail Surgery
Please handle ALL of the surgical instruments used in this lab with
extreme care, especially the very fine Vanna scissors and Dumont
forceps. These are expensive, irreplaceable, and will
likely be irreparably damaged if they are dropped or allowed to
rust.
The tiny Minuten pins took
some time and effort to manufacture "in house". Try NOT to
lose them. Keep them imbedded in the Sylgard of the
chamber when not in use, and rinse the chamber after each use.
1)
Choose either a Lymnaea or Helisoma snail.
2) Watch the
appropriate dissection video at least twice. The surgical
procedures for the two species are similar, but not identical
and conform loosely to the sequence below.
3) Anesthetize the
snail by placing it in 5% EtOH or 10% Listerine for 5-10 minutes.
4) Carefully peel back
and chip away the shell from the head of the snail using iris
scissors, fine forceps, and/or your thumbnails. Continue
until the head and mantle are completely exposed.
5) Use straight pins to
pin out the snail in a large Sylgard-lined Petri dish, filled
with standard Snail Ringer's. Make sure that the snail is
pinned ventral side down. Place one pin though the
center of the body and visceral mass posterior to the mantle,
then one pin through the head to either side of the antennae.
Fold the mantle back from the head and pin it on each side.
6) Using the smallest
iris (Vanna) scissors, make a shallow midline dorsal cut through
the mantle and up to the anterior end of the head. Reflect
the outer body wall back and pin it to each side.
7) Remove the overlying
reproductive organs.
8) Locate the midline
upper digestive system, with the multi-lobed brain
wrapped around the esophagus and the anterior buccal mass.
The easiest way to identify the brain is by following the white
fibrous connectives leading into it from all sides.
9) Cut through the
esophagus posterior to the brain, carefully pull it out
through the brain, and pin it out in front, so that the ventral
surface of the buccal mass is now exposed.
10) Locate the tiny two-lobed
buccal ganglion on the exposed ventral surface on the buccal mass.
Again, the easiest way to find the ganglion is by following the
white fibrous connectives leading into it from all sides.
11) Prepare one of
the small plastic Petri recording dishes by filling it with ~1mm of
snail
Ringer's.
12) Use the fine Vanna scissors to
carefully detach the buccal
ganglion by cutting through each connective, as far from the
ganglion as possible. Leaving long connectives
attached will make the ganglion easier to handle, make it easier
to pin out, and minimize damage to the neurons within the
ganglion. Using the finest Dumont forceps, transfer the ganglion
to the recording dish, grasping only the connectives and not
the ganglion itself.
13) Relocate the brain and cut
through the dorsal commissure. This will "unfold" the
brain and make it easier to see the multiple lobes in each
hemisphere. Carefully detach the brain and transfer it to
the recording dish, again cutting each connective as long as
possible and handling only the connectives.
14) Use the tiny Minutin pins in
the recording dish to pin out both the brain and the buccal
ganglion dorsal side up. Try to pin both out near
the center of the dish, so that they will be more accessible
when it comes time to record from them with microelectrodes.
B.
Desheathing
the Ganglia with Protease
1)
Use a plastic pipet to remove the Ringer's from the two ganglia
and immediately cover each with a droplet of protease solution.
Leave the protease on the buccal ganglion for ~1 minute and on
the brain for ~2 minutes. When you remove the protease,
immediately recover each ganglion with fresh Ringer's.
Do not leave either ganglion exposed to the air for more than
20 seconds. If you leave the protease on two long the
ganglia will simply turn to mush and be unusable.
2) When you are
finished with the protease treatment, carefully rinse
both ganglia by replacing the Ringer's at least twice more, then
add enough Ringer's to refill the dish, so that the ganglia are
covered.
3) If necessary, re-pin
each ganglia so that it is stretched and relatively immobile.
If the ganglia can move, then impaling cells with a
microelectrode will be much more difficult.
C.
Recording
Setup
1)
Pull at least four microelectrodes on the Sutter Instruments
puller, using the program specified by the instructor.
Fill and mount one in the wire-core half-cell holder and at
least one in a pellet half-cell holder.
Note:
Use 0.6 M
K2SO4 to fill the electrodes.
2) Turn on the
fiber-optic microscope lights, the PC, the function generator,
the audio amplifier, and the PowerLab.
3) Set the Audio
Selector Switch to VCO and the audio amplifier to TUNER with the
MONO switch out. Adjust the function generator and the
audio amplifier to produce a pleasant tone at a low volume.
4) Turn on the Model
1600 Neuroprobe amplifier with the switch on the back.
Confirm that the OUTPUTS are connected as follows: X1 to Powerlab Ch1,
X10 to the function generator VCO input (on the
back), and Current to PowerLab Ch 2. Confirm that the
CURRENT INJECTION TRANSIENT inner knob is turned all the way
counter-clockwise (off), both CAP COMP knobs are turned
all the way counter-clockwise (minimal), and all of the black
push-buttons (except METER OFF) are in the out (off) position.
5) Start Scope.
Set Input A to Ch 1 and Input B to Ch 2. Set both input
channels for single-ended recording with no filters. Set
Ch A to 200mV range and Ch B to 100mV range. Set the time
base to 200mesc and maximal sampling rate. Set Sampling to
Repetitive. Adjust the screen display so that Ch A
occupies about 2/3 of the vertical display area. At
this point, you may want to save a Scope Settings file named
"Snail Brain Settings" to the desktop, to make setup faster for
future sessions.
Note: Input A (Ch1) and the
audio tone will monitor the voltage at the microelectrode,
exactly as in the previous muscle RP lab. Input B (Ch2)
will monitor the current injected through the microelectrode by
the Neuroprobe amplifier. The display scale for CH2 is
10mV per 1nA injected current.
6) Check that all of
the ground and shield cables within the cage and between the
cage and the PowerLab and amplifiers are appropriately
connected.
D.
Electrode
Testing and Compensating
1)
Position either a Petri dish or a ganglion chamber filled with
Ringer's in the center of the microscope stage. If
necessary, secure it in place with a few small pieces of clay.
2) Mount an electrode
and holder to the headstage in the micromanipulator.
Position the electrode tip in the Ringer's. Make sure
that either the headstage ground pellet, or the cage shield
ground pellet is in the Ringer's bath.
3) To zero, impedance
test, capacitively compensate, and current compensate your
microelectrode, follow the entire Set-Up Procedure on pages 9
and 10 of the Neuroprobe Amplifier Manual, up to step #20.
Only use electrodes with a tip impedance between 10 and 60 Mohms.
4) Congratulations!
You are now ready to record from your snail brains for the
remaining five minutes of the lab session (hopefully more).
Save sufficient traces to complete Data Sheet Item # 1 first to demonstrate this.
|