I.
GATED EQUIVALENT CIRCUIT MODEL - EXPERIMENT
A. Initial Setup
1)
Start up the PC (if necessary), turn on the PowerLab box,
and open the AP Template file. This will launch the Scope
application, make appropriate PowerLab settings, and load a set
of sample traces.
2)
Examine the PowerLab display, which should appear similar to the
sample in the link to the right. The upper trace displays
Input A. A "template" action potential which you are to attempt
to match appears in red. Additional lines correspond to
the equilibrium potentials for sodium (ENa),
potassium (EK),
and chloride (ECl),
as well as the resting potential (VR)
and the threshold potential (Vth).
The lower trace displays Input B. The output pulse of the
PowerLab stimulator appears in blue.
3)
Under the Setup / Sampling menu verify that Mode is set to
Repetitve.
4)
Connect two screws on the gated equivalent circuit model (GECM)
board to the CH1 input of the PowerLab using a BNC-to-double
banana plug cable and two alligator clip adapters. Make sure
that the "IN(SIDE)" screw on the GECM is connected to the active
(nontab) banana plug and the "OUT(SIDE)" screw on the GECM is
connected to the common or ground (tab) banana plug.
5) Plug
all three power supplies into the GECM, matching them by color (green,
red, and yellow).
6) Turn
on the power strip(s) for the power supplies.
7)
Connect the PowerLab Output+ to the CH2 input via a BNC cable
and T-connector. Connect the free end of the T-connector to the
white stimulus input on the GECM via a BNC cable,
BNC-phono plug adapter, and phono plug to 1/8" phone plug
adapter.
8) Set
all three switches on the GECM to the off (down) position.
9) Start
a continuous series of PowerLab sweeps via the Start button on
the Scope display. The new red membrane potential trace should
show a constant value corresponding to the resting potential VR of
about -70 mV, with a rounded upwards deflection at the time of
the stimulus pulse. The new blue stimulus trace should match the
template stimulus trace with a square upwards deflection. You
should also hear a distinct "click-click" produced by a magnetic
relay on the GECM board. If these initial conditions are not met,
consult with the instructor.
10)
Disconnect the stimulus cable from the GECM.
B. A "Blind" Experiment
You have
three toggle switches and one push-button switch which you can
manipulate on the GECM. These are labeled A, B, C, and D. Your
task for this part of the laboratory exercise is fairly simple:
1)
Experiment with throwing switches A, B, and C and pushing button
D as Scope repetitively sweeps out new traces on the computer
screen.
2)
Continue until you have determined a set of starting switch
positions, a sequence of switch changes, and a timing of switch
changes which results in a trace closely matching the template
action potential. All switches should be returned to their
initial positions by the end of the trace.
3) Once
you think that you have the correct timing, reconnect the
stimulus cable to the GECM via the 1/8 inch phone plug adapter.
Now the Scope software will produce the stimulus pulse and you
just have the three toggle switches to worry about.
Q1:
When you think you have a solution which closely matches the
template, call the
instructor over to demonstrate your solution.
If you are
having trouble figuring out how accomplish this, here is a
systematic way to proceed:
Step 1:
Determine the effect of each switch on the membrane potential
and the time-course of
each such effect. For each toggle switch up (towards the
circuit board) is on and down (away from the circuit board) is
off. For the push-button switch the depressed position is on
and the released position is off. Note: Not
all of the switches have independent effects - the effect of
turning on one switch may depend on whether another switch is on
or off.
Step 2:
Break the action potential template into components. Determine
what switch and/or button manipulations are required to produce
each component. For example:
What switch setting(s) produce(s) the resting
potential VR of
about -70mV?
What produces the initial stimulus hump (at
5.0-5.7 seconds)?
What produces the rising, depolarizing phase
(at 5.7-6.0 seconds)?
What produces the falling, repolarizing
phase(at 6-7 seconds) and
undershoot, hyperpolarization
phase (at 7-8 seconds)?
What produces the recovery from undershoot (at
8-11 seconds)?
How and when during the course of the action
potential can each switch be
restored to its starting position?
Step 3:
When you have duplicated the rough shape of the action potential
template, work on refining your timing to provide a closer
match. Refine your technique to get the simplest series of
switch manipulations which reliably produces an action potential
template match.
Stop
reading here until you have successfully completed sections A
and B.
II.
GATED EQUIVALENT CIRCUIT MODEL - COMPONENTS
A. The Membrane Equivalent Circuit
Examine
the circuit diagram for the GECM with which you have been
experimenting, by clicking on this link. Work with
this diagram until you thoroughly understand what each component
represents in an actual neuron membrane. You should be able to
identify each of the electronic components on the circuit
board. We will be exploring the equivalent circuit model at
length in class, so bring your questions with you.
The GECM
model has six small circuits, arranged in parallel, which model
specific components of the neuronal membrane - specifically a
space-clamped axon. Working from left
to right in this diagram and on the GECM circuit board:
1)
Recording - This consists of the PowerLab amplifier,
connected across the modeled neuronal membrane (the two screws
labeled “OUT” and “IN” at the extreme right end of the GECM
board). By convention, transmembrane potentials are recorded
inside relative to outside, hence the active (positive) input of
the PowerLab is connected to the inside of the simulated neuron
and the common (negative, ground) input to the outside. A 40 MW
series input impedance, produced by four 10 MW
resistors (brown-black-blue bands; diagonally mounted at the
lower left end of the circuit board in the active recording
lead), functions as a voltage divider. This insures that the
PowerLab drains negligible current off the rest of the circuit
and does not itself affect the measured membrane potential. It
also scales down the measured voltage into the appropriate
millivolt range. An additional recalibration of the recorded
potential is performed by the PowerLab Scope software.
2)
Sodium - According to Ohm's Law for ionic currents,
sodium is pushed across the membrane by a driving force equal to
the membrane potential minus the sodium equilibrium potential (Vm -
ENa). In this model, ENa is set to approximately +55 mV by a +4.5 V voltage
supply (green phone jack). Sodium current flows through
selective membrane channels, represented in this model by two
routes:
a)
The voltage-gated, active sodium conductance is
represented by a 6.8 KW
resistor (blue-gray-red bands) wired in series with the first
two switches, labeled "A" and "B", on the board.
These two switches, marked "A" and "I" in the diagram above,
represent the sodium "activation" and "inactivation" gates of
the membrane sodium channels. Both switches must simultaneously
be in the "on" position in order for current to flow through
this low conductance route.
b)
The resting sodium conductance through these same
channels is represented by a separate route through a 470 KW
resistor (yellow-purple-yellow bands).
Current flowing through the combined sodium conductance pulls
the measured membrane potential towards the sodium equilibrium
potential ENa
(approx. +
55 mV).
3)
Potassium - Potassium is pushed across the membrane by a driving
force equal to the membrane potential minus the potassium
equilibrium potential (Vm -
EK). In this model, EK is set to approximately -90 mV by -9.0 V voltage
supply (red phone jack). Potassium current also flows through
selective membrane channels, represented in this model by two
routes:
a)
The voltage-gated, active potassium conductance is
represented by a 22 KW
resistor (red-red-orange bands) wired in series with the third
switch, labeled "C" on the board. This switch, marked
"A" in the diagram above, represents the potassium "activation"
gates of the active membrane potassium channels. This switch
must be in the "on" position in order for current to flow
through this low conductance route.
b)
The resting potassium conductance through these same
channels is represented by a separate route through a 100 KW
resistor (brown-black-yellow bands).
Current flowing through the combined potassium conductance
routes pulls the measured membrane potential towards the
potassium equilibrium potential EK
(approx. -90 mV).
4)
Chloride - Chloride is pushed across the membrane by a
driving force equal to the membrane potential minus the chloride
equilibrium potential (Vm -
ECl). In this model, ECl is set to approximately -70 mV by a -6.0 V voltage
supply (yellow phone jack). Chloride current can flow through
the membrane only through a small constant chloride conductance,
represented by a 1 MW
resistor (brown-black-green bands). This voltage-gated
conductance pulls the measured membrane potential towards the
chloride equilibrium potential ECl
(approx. -70 mV) which is approximately equal to
the resting potential VR.
In your particular model, ECl
may be anywhere in the Vrest
+ or - 5mV range
5)
Membrane Capacitance - As discussed in earlier labs, the
lipid bilayer is a nonconductor situated between two conductive
mediums, the intracellular and extracellular spaces. Because of
this, dissimilar charges can accumulate on the two sides of the
membrane, held loosely in place by their electrostatic
attraction across the membrane. This property of separating
charge is called capacitance. The membrane capacitance for this
model is set to 30 mF
and is produced by three 10
mF
capacitors (blue cylinders) wired in parallel. This capacitance
value was chosen to produce model kinetics (rates of potential
change) which are about 1000 times slower than those of an
actual neuronal membrane. These capacitors "round out" the
shape of the action potential. If these capacitors are removed
from the circuit any change in membrane conductance produced by
throwing a switch results in an instantaneous membrane voltage
change.
6)
Stimulation - Action potentials are normally triggered by
transient depolarizations of the neuronal membrane. In this
model stimulus depolarizations are produced by closing either
one of two switches which temporarily short the inside and
outside of the membrane together through a 68 KW
resistor (blue-gray-orange bands) located at the upper right of
the circuit board. This actually corresponds well to an
excitatory postsynaptic potential produced by opening up a
non-selective monovalent cation channel, such as that associated
with the nicotinic acetylcholine receptor or with the fast
glutamate receptor. In the model this stimulating circuit may
be manually activated via the red push-button switch labeled "D"
or automatically via the magnetic relay which is controlled by
the PowerLab stimulator output (white phone jack).
B. Voltage-Gated Conductances
In this
simulation you control the switch "gates", which in turn control
the relative conductances to sodium and potassium and produces
the current fluxes and membrane potential changes associated
with the action potential. However, in an actual neuron, these
gates are controlled by the membrane potential itself, in
specific time-dependent manners. Before you leave this
simulation it is important that you reconcile your AP solution
with the following "rules" for gate (switch) states and
kinetics.
Note:
The terminology for switches vs. gates is confusing - an ion
channel gate conducts when it is open while an
electrical switch conducts when it is closed.
Thus closing a switch to the "on" position models opening the
corresponding gate, while opening a switch to the "off" position
models closing the corresponding gate. For the discussion below
we will use "on" and "off" to refer to switch states and
"open" and "closed" to refer to the corresponding gate
states.
1)
Sodium Activation - At rest this switch should be off,
(deactivated and closed gate) blocking conductance through this
route. It turns on (activates and opens) rapidly
in response to a suprathreshold depolarization, such as that
provided by the stimulus pulse. It turns off
(deactivates and recloses) rapidly when the membrane
repolarizes towards the resting potential or below.
2)
Sodium Inactivation - At rest this switch should be on (deinactivated
and open gate). It turns off (inactivates and closes)
slowly as the membrane depolarizes. It turns on (deinactivates
and reopens) slowly as the membrane repolarizes back
toward the resting potential or below. Very Important Note:
Although deactivation and inactivation will both
close a sodium channel, they are NOT the same thing.
Deactivation and inactivation refer to two separate processes
governed by two structurally separate gates with very different
voltage dependencies and kinetics (speeds of action).
3)
Potassium Activation - At rest this switch should be turned
off (deactivated and closed gate. It turns on
(activates and opens) slowly as the membrane
depolarizes. It turns off (deactivates and recloses)
slowly as the membrane repolarizes back toward the resting
potential or below.
At this
time you should rerun and perhaps modify your action potential
solution so that it conforms to these rules.
Q2:
Demonstrate your new solution to the instructor.
|