I.
EQUIPMENT SETUP
We will be using the electronic stimulator to initiate muscle
contractions by delivering pulses through a pair of silver wire
electrodes mounted in a probe. Contractions of the muscle will
be conveyed via a thread to the central rod of a linear
displacement transducer. This transducer converts changes in
muscle length to changes in electrical resistance. A bridge
amplifier converts theses resistance changes to voltage changes.
The output cable of the bridge amplifier carries this voltage
signal to a PowerLab input. Suspending a fixed weight to the
transducer rod allows for isotonic muscle contraction under a
fixed load. |
The
electronic equipment needed for this experiment includes a
displacement transducer, an ETH-400 transducer bridge, an
electronic stimulator, a stimulator probe with silver wire
electrodes, and a PowerLab/PC station.
To save
time in this lab you should split your group into two subgroups,
one of which will begin perform the dissection (Section II)
while the other sets up the equipment as described below.
Equipment Setup Procedure:
1) Clamp
the body of the displacement transducer about halfway up a ring
stand. The "hat" should be on top. Connect the transducer
cable to Channel 1 of the ETH-400 bridge amplifier. Connect the
Channel 1 output of the ETH-400 to the + CH 1 input jack of the
PowerLab box, using a BNC cable. Make sure that the bridge
power cable is plugged in and that the red light on the bridge
is on.
2) Mount
the femur clamp about 3/4 of the way up the same ring stand.
Mount the stimulator probe just above the femur clamp.
Carefully plug the stimulating electrodes into the stimulator
probe. Attach the probe leads to the electronic stimulator
outputs, red-to-red and black-to-black via the adapter cable.
Use a double-banana to BNC cable to also connect the stimulator
cable to the + CH 2 input on the PowerLab, so that you can
monitor the stimulator output directly.
3)
Connect a BNC cable from the + Output jack of the PowerLab to
the SYNC TRIGGER IN jack of the stimulator (near the lower left
of the front panel). This cable will allow the internal
PowerLab stimulator to trigger the SD9 electronic stimulator.
4) Check
out all of the cable connections with the instructor before
proceeding.
5) Make
sure that the stimulator MODE is set to OFF. Turn
on the stimulator, turn on the PowerLab box, boot the PC, and
launch Scope.
6) Set
Input A to Ch1 and Input B to Ch2.
Adjust the display area so that most of the vertical space is
occupied by Input A. Set Samples:2560 and
Time:500 msec in the submenu at the bottom of the Speed
menu.
7) Under
the Display. . . menu chose Overlay Stimulator and
set it to display the PowerLab stimulator trace at the top of
the display.
8) Under
the Setup:Sampling menu set Mode:Single and
Source:User. Under the Setup:Stimulator menu set
Mode:Pulse, Delay to 100 msec, Duration
to 5.0 msec, and Amplitude to ~4 volts.
9) Use
the Input A Input Amplifier... to set the display range
to 1 V and activate the Positive input and the Invert.
Use the Input B Input Amplifier... to set the display
range to 10 V and again activate only the Positive input.
10) Use
the Units Conversion... option on Channel 1 to
have Scope convert directly from mV to mm muscle
contraction so that the vertical axis reads in mm, rather than
mV. The conversion ratio is 1.2V to 1.0cm. (see the instructor
for help, if necessary).
11) Double
check the setup of all of your equipment, and make sure that you
know the locations and functions of all of the controls. Refer
to the instrumentation tutorial or consult the instructor if you
have any questions.
II.
DISSECTION AND MOUNTING OF THE NERVE/MUSCLE PREPARATION
As in the previous experiment, try to work quickly and
carefully. Wear gloves. Read all the way through the dissection
procedure before you start. Handle both the nerve and the muscle
carefully, and touch them only with the glass rod and hook. If
you stretch the muscle, you will reduce its contractility, so
try not to stretch it during the dissection. Don't let the
muscle or nerve touch the outside of the frog's skin. Keep all
exposed tissue moist with frog Ringer's. |
Dissection Procedure:
1) Put
on a pair of gloves. Obtain yet another decapitated frog from
your instructor or one of the other groups. Try not to ruminate
on the fact that amphibians are in a serious decline worldwide.
Cut off a hind leg at the upper end of the thigh and, if there
is still one leg left, give the rest of the frog to one of the
other laboratory groups.
2)
Remove the skin completely from the leg. Try not to let the
outside of the skin touch the muscle of the leg. Wash your
hands and change your gloves.
3)
Expose the sciatic nerve in the thigh, free it between the knee
and the hip, and tie a thread around it as close as possible to the hip.
If you have correctly identified the sciatic nerve the leg
muscles should twitch violently when you tighten the knot on the
tread. Sever
the nerve proximal (towards the body of the frog) to the
tie, then lay the nerve along the gastrocnemius muscle.
Wet the nerve and muscle with a few drops of Ringer's solution. Be sure
not to touch the nerve with your fingers or with any metal
instruments, only with the thread and the glass hooks.
4)
Carefully cut away the thigh muscles (but not the nerve!)
leaving the femur (thigh bone) exposed. Use a needle to poke a
hole in the Achilles tendon at the heel. Tie a thread through
the hole in the tendon, then use a sharp scalpel or scissors to
separate the tendon from the ankle distal to the thread.
5)
Gently lift the gastrocnemius and sciatic nerve with a glass
rod, while cutting through the tibiofibula (lower leg bone) and
the other lower leg muscles. Make your cut below and close to
the knee. You will end up with the gastrocnemius muscle
attached to the femur at one end and the thread at the other
end, with the muscle's nerve supply intact.
6) Place
the shaft of the femur between the jaws of the femur clamp and
tighten the jaws securely. Allow the nerve to hang along the
muscle for now, and moisten both with Ringer's. Do not let
the nerve touch the femur clamp or the ring stand. Tie the
Achilles tendon thread to the upper end of the transducer rod,
just above the plastic "cap". The ~6 gram weight of the
transducer rod will function to keep a slight tension on the
muscle.
7) Now
carefully adjust the position of the clamp, transducer,
and/or bridge offset knob until the muscle is perpendicular to
the femur, the transducer rod is vertical, the thread is
vertical, and Input A reads 0 volts. Using the running
Input Amplifier... window for Input A will help
you with this last determination.
8)
Position the silver stimulating electrodes on the probe tip
as close to the knee joint as possible. Gently wrap the
nerve around the stimulating electrodes and moisten it with
Ringer's solution.
9) At
this point you should ask the instructor to come over and do a
quick stimulation check to make sure that your nerve/muscle
preparation is viable and responsive. Once you have
confirmed this, rewet the nerve with Ringers, then coat all of the exposed nerve with
a layer of Vaseline to keep it from drying out.
Be sure
that the muscle and exposed sciatic nerve are kept moist with
Ringer's solution.
DO NOT ALLOW RINGER'S SOLUTION TO COME IN CONTACT WITH
ANY OF THE RECORDING EQUIPMENT, INCLUDING THE TRANSDUCER.
Don't allow Ringer's solution to accumulate on the transducer
cap. The viability of the preparation is limited and it will
begin to deteriorate as soon as the animal is sacrificed.
Therefore, it is important to avoid unnecessary stimulation of
the muscle and to be efficient in your use of time. If the
muscle gets "twitchy" and starts contracting on its own, your
best bet is to make sure that the nerve is not being stretched,
wet down both the nerve and muscle with Ringer's solution, and
simply wait for the preparation to "calm down".
III.
RECORDING FROM THE NERVE/MUSCLE PREPARATION
Skeletal muscle contraction is generally studied under one of
two experimentally imposed conditions: isometric (constant
length) or isotonic (constant load). We will be studying the
gastrocnemius muscle under isotonic conditions, where the muscle
contraction moves the transducer lever, lifting a set weight.
Under these circumstances, the muscle does external work by
displacing the weight. The degree of shortening of the entire
muscle (contraction) is reflected by the degree of displacement
of the transducer rod. |
As you
work through the following procedures, be sure to save important
Scope and Chart records to disk. Keep track of
your chart records by making Comments. Save your file
periodically. Rename your file periodically using the Save
As... option, so that older versions of your file are
maintained (so if you mess up you don't lose all of your
data). When you are instructed to print out a Scope or
Chart record, do that after the experiment, while
you are preparing your data sheet. Use your time during the
experiment to collect the data, making sure you have obtained
and saved the records you will need for the print-outs and
measurements you will make later.
Again,
keep the preparation moistened with Ringer's.
A.
Components of the Muscle Twitch:
1)
Using a stimulus duration of 0.1 to 0.5 msec. and an
electronic stimulator delay of ~0 msec, deliver pulses of
gradually increasing voltage by repeatedly triggering Scope
sweeps, while you watch the muscle and the Scope trace.
When you reach a stimulus voltage that elicits a visible twitch,
you may want to clear out the earlier traces, using New
under the File menu.
2)
Continue
to increase the voltage until a maximal twitch is produced. Be
sure to use Page Comments liberally to label each section
of the record with the stimulus parameters.
3)
Now look back through the records of the twitches that were
produced as you increased the stimulus voltage. Pick out a
record for which the twitch is roughly one-half-maximal, and one
where it is roughly one-tenth-maximal. Stimulate the muscle
again if you need to get a closer approximation to either of
these response amplitudes. |