This lab, and several subsequent labs, will involve
penetrating individual cells with fine-tipped glass
microelectrodes and measuring electrical potentials across the
cell membrane. This general set of techniques is called
"conventional" intracellular recording.
The first goal of this lab is to become proficient at
manufacturing, handling, and using glass pipette
microelectrodes, penetrating cells, and using the Neuroprobe
amplifier to measure trans-membrane potentials. The second goal
is to use these new skills to study the variability in resting
potentials in crayfish muscle cells and the dependence of the
resting potential on extracellular ionic concentrations. The
third goal is to verify the RC equivalent circuit model as a
good representation of the passive response properties of a
membrane to current injection.
Although specialized instruments and techniques have been
designed to facilitate intracellular recording, it remains MUCH
more demanding than extracellular recording. The next several
labs will all depend on and build on the skills you will acquire
in this lab, so you really need to keep trying until you
understand both theory and practice and can make it work. This
sequence of Crawdad Workshop labs may well be the most
painstaking, infuriating, character-building, and/or satisfying
work you will ever perform in a laboratory. Enjoy them while
they last. |
|
|
I. PREPARING FOR INTRACELLULAR RECORDING VIA GLASS
MICROELECTRODES
The good news for conventional intracellular recording is that
transmembrane potentials are on the order of tens of millivolts,
or 100 to 1000 times larger than the extracellularly recorded
action potentials of Crawdad Lab #2. This makes environmental
electronic noise less of a problem. Typical amplification
levels are x1 or x10, rather than the x1,000 or x10,000 that you
are used to. Also, the resting potentials in this exercise
are VERY slowly changing, so high-frequency noise, including 60
Hz line noise, can be filtered out.
Unfortunately, there is a wealth of bad news associated with
intracellular recording. First, penetrating an individual
cell requires much finer electrode positioning than does simply
sucking up an entire nerve root. Second, intracellular
recordings must be "DC"; because slowly changing potentials are
an important part of the signal you don't get to filter out slow
baseline drifts. This requires both excellent mechanical
isolation from room vibrations and a special kind of amplifier.
Third, the electrode must be able to penetrate the cell without
significantly damaging it. To do this you must use glass
microelectrodes with extremely fine and sharp tips. These
electrodes are fragile, finicky, fickle, frustrating, and
unforgiving, and must be handled and positioned with extreme
care. Finally, the glass microelectrodes themselves
introduce distortions into the recorded signal, and these
distortions must be compensated for by unique amplifier
adjustments for each new electrode.
As described in Appendix D of the Crawdad manual, glass pipette
microelectrodes are manufactured by heating and drawing a glass
capillary tube. As you discovered in the first lab, we
have an "electrode puller" which does this in a semi-automated
fashion and produces fairly consistent results.
Unfortunately, because glass is actually a liquid and will flow
over time, microelectrodes have to be used within a few hours of
when they are manufactured.
Electrical contact will be made with the interior of a
penetrated cell by filling the interior of the microelectrode
with a salt solution (generally 3M KCl). The electrode is
mounted to an electrode holder, where this internal KCl solution
is in contact with a chlorided silver (Ag/AgCl) pellet. The
pellet is, in turn, wired directly to the live recording lead.
Thus, when a cell is penetrated by the microelectrode tip, its
internal cytoplasm is in electrical contact with the live
amplifier input via this KCl solution and Ag/AgCl pellet. The
extracellular reference electrode for intracellular recording is
generally a Ag/AgCl pellet in the surrounding bath solution. Why
Ag/AgCl? For DC recording an electrically stable interface must
be created between both the intracellular and extracellular
fluids and the recording lead. Simple metal surfaces immersed in
a saline solution create DC "junction potentials" - essentially
they act like miniature batteries whose current-passing
properties slowly change as ions are electrolytically plated
onto their surfaces. A silver pellet pre-plated with a layer of
silver chloride produces a stable surface which is largely free
from these problems.
If, in the process of recording, the amplifier circuit were to
draw appreciable current out of the cell, it would directly
affect the local membrane potential and interfere with accurate
recording. This problem is solved by placing a "high-impedance"
headstage (essentially a large voltage divider - see lab #3)
between the cell and the amplifier, so very, very little current
is drawn off.
In order to penetrate the cell and make a tight electrical seal,
the microelectrode tip must be smooth and have a very small
diameter The relative diameter of the tip can be easily
determined by measuring the impedance (resistance to current
flow into the tip) of the electrode. Low impedances
indicate broken electrode tips that are two large and rough to
penetrate cells. Excessively high impedances indicate
electrodes that are internally blocked by debris or air bubbles.
Such electrodes will generate too much recording noise;
essentially they act more like antennae than electrodes.
A final problem is that at the tip of the microelectrode two
conductive salt solutions are separated by a very thin layer of
glass. This makes the tip behave like a capacitor which can
dramatically attenuate or eliminate high-frequency (rapidly
changing) components of the signal. Fortunately the amplifier
has a "capacity compensation" circuit specifically designed to
compensate for this effect - sort of like the "Dolby" system on
stereo tape decks. |
A.
Preparing Microelectrodes
Handle glass microelectrodes with EXTREME CARE and always
DISPOSE OF THEM properly in a SHARPS CONTAINER. Glass
microelectrodes have sub-microscopic tips and are EXTREMELY
sharp and fragile. Careless handling will result in you or
someone around you becoming hurt. Unattended microelectrodes
invariably find their way into human bodies, often finger
joints, with unpleasant and untreatable results.
1) Pull
at least 6 microelectrodes from thin-walled 1.0 mm
capillary tubes, using the skills you practiced in lab #1. Note
that each capillary tube will provide two electrodes. Handle
the capillary tubes carefully, both before and after pulling.
Avoid touching the center region of the tube with your bare
fingers, the oil will interfere with proper heating and drawing
of the glass and the sharp tips will break off in your fingers.
Glass microelectrodes can be stored wet or dry for up to 24
hours by pressing them into clay strips in a covered petri dish.
2) To
prepare a microelectrode for use, first stand it tip up
in a small beaker of 3M KCl for at least 5 minutes. Capillary
action will draw the KCl solution up to the tip and fill the
tapered part.
3) After
5 minutes, carefully backfill each microelectrode with 3M KCl
using a 5ml syringe and a microfil fiber needle. Backfill the
micropipette slowly, withdrawing the needle tip as you go. This
will minimize bubbles and prevent blowing the tip off of the
micropipette. Small bubbles along the thick barrel of the
microelectrode will not interfere with recording, but avoid
bubbles near the tapered tip.
4)
Loosen the screw cap of a half-cell electrode holder and
completely fill it with 3M KCl, being careful to leave no
bubbles inside. VERY GENTLY push the barrel end of a
filled microelectrode into the electrode holder and through the
internal rubber seal. If you apply too much pressure to the
electrode it could snap and become a permanent part of your
hand. Carefully screw the cap in - this only needs to be tight
enough to hold the microelectrode without leaking.
OVER-TIGHTENING THE SCREW CAP CAN SHATTER THE ELECTRODE AND/OR
RUIN THE HOLDER.
5) Make
sure that the Model 2700 Microelectrode R/C Meter is turned on,
that the test beaker is filled with crayfish Ringers, and that
the ground electrode is suspended in the beaker. Mount the
electrode holder into the testing lead, then rotate the clamp
arm down so that the tip of the electrode is immersed in the
Ringers. Set the RANGE knob to 20 MW.
Rotate the CAPACITY ADJ. knob fully counter-clockwise. Set the
DISPLAY switch to MEASURE. The meter should now be displaying
the electrode tip impedance (or resistance) and capacitance.
Change the RANGE if the meter directs you to do so. A more
accurate reading of the tip impedance can be obtained by turning
the CAPACITY ADJ. knob clockwise unit the meter reads "Over
Compensated", then backing it off slightly.
6) An
ideal electrode should have a tip impedance of 0.5-2.0 Mohm.
When you get a good electrode, carefully remove the holder from
the meter and transfer it to the high-impedance headstage in
your recording rig. Don't remove the electrode from the
holder. Handle the electrode and holder carefully; any
contact with the tip of the microelectrode will either break it,
clog it, or imbed it in your flesh, making it useless for
recording.
7) If
the tip impedance is below .5 Mohm,
then the tip has probably been broken. If the impedance is
above 10 Mohm,
then the tip is probably clogged. In either case, loosen the
screw cap on the holder, carefully remove the microelectrode,
discard the microelectrode into a Sharps container, and begin
again with Step 3.
8) Steps
3-6 must be performed individually on each electrode before it
can be used. In practice it will be a good idea to have one
electrode in use in your recording rig, and one prepared as a
backup at all times. You have two half-cell electrode holders
and at least two members in your group for exactly this reason.
B.
Neuroprobe Amplifier Setup
1) Turn
on the power to the Model 1600 Neuroprobe Amplifier. There are
two power switches, one on the back and one on the front. Set
the METER to OFF, the HIGH-RANGE to OFF (out), all of the
CALIBRATE buttons of OFF (out), the CAPACITY COMP. knob fully
counter-clockwise, the CURRENT INJECTION switch to OFF, all
three CURRENT INJECTION KNOBS to fully counter-clockwise, the DC
OFFSET switch to OFF, the DC OFFSET knob fully
counter-clockwise, and the LOW-PASS filter knob to OFF.
2) Allow
the Neuroprobe amplifier to warm up for 5 minutes. To verify
that the power is on you can briefly depress the METER PROBE
button. This should activate the LED display window. Set the
METER back to OFF while the amplifier warms up.
3) Turn
on the microscope and fiber-optic lights. Fill an extra Sylgard-lined
recording dish with about 1 cm of cold crayfish Ringers.
Carefully position the microelectrode tip so that it is barely
immersed in the Ringers bath. Place the reference/ground pellet
in the bath.
4)
Depress the METER PROBE button. Adjust the DC OFFSET switch and
knob until the meter reads 000. This eliminates any constant
"DC" potential between the active microelectrode and the
reference bath pellet.
C.
Audio Monitoring
1) Make
sure that the x10 output of the Neuroprobe amplifier is
connected to the VCO input on the back of the function
generator.
2)
Switch the audio monitor switch to VCO.
3) Turn
on the power to the function generator. Set the function
generator to sinusoidal output in the 100Hz range. Turn the
amplitude all the way down (counterclockwise).
4) Turn
on the stereo audio amplifier and select Tuner. Adjust the
audio amplifier volume and/or the function generator amplitude
to produce a steady tone from the right speaker.
D.
PowerLab Setup and Calibration Practice
1) Turn
on the PC and PowerLab box and start up Scope. Make sure that
the X1 output of the Neuroprobe amplifier is connected to the
CH1 + input of the PowerLab.
2) Turn
off Input B. Set Input A to 100 mV, and Positive. DO NOT
activate either the AC, 50 HZ LP, or line filters. Set the
Time Base to 20msec and 1024 samples (40 kHz). Open the
Display; Computed Functions . . . window and set the display for
Channel A Only. Under the Setup; Sampling menu choose
Repetitive.
3) Click
on the Scope Start button to begin monitoring the Neuroprobe
amplifier output. Open the Input Amplifier... and Display
Offset... windows to confirm that Channel A is reading 0 volts.
Adjust the Neuroprobe DC OFFSET knob to confirm that the
Neuropobe and PowerLab displays register the same values (within
2-3 mV). Return the DC OFFSET to 000, then close the Display
Offset... and Input Amplifier... windows on Scope.
4)
Depress the HIGH RANGE and ELEC TEST buttons on the Neuroprobe.
This passes a 100 Hz x 10 nA square wave signal through the
electrode. The Neuroprobe LED meter will show the electrode
impedance in Mohm.
The amplitude of the signal on the PowerLab will be 10 mV/Mohm.
In other words, a 50 mV square wave on the Scope display
corresponds to a 5 Mohm
electrode resistance.
Q1:
These electrode test current, voltage, and resistance values
should conform to Ohm's Law. Do they?
5)
Carefully adjust the CAPACITY COMP knobs clockwise to
"square-up" the test signal display on Scope. (NOTE: For some
bizarre reason the inner knob is the coarse and the outer knob
is the fine control.) This compensates for the effects of the
tip capacitance. Stop just short of the point where capacitive
"spikes" occur at the square-wave corners and the signal becomes
unstable or "rings”. If CAPACITY COMP is set too high, or if
the OVERRIDE button is depressed, the resulting feedback will
"buzz" the electrode tip. This is handy for breaking into
cells, but prolonged buzzing is bad for both the electrode tip
and the amplifier. Briefly press the OVERRIDE button to see the
effect on the Scope display. |
|
|
Data Sheet Item #1:
Print out a set of three labeled
Scope traces illustrating an electrode test with capacitively
under-compensated, ideally-compensated, and over-compensated
settings. Indicate on the ideally-compensated trace the
electrode resistance and how you measured it.
|
|
|
6)
Release both the ELEC TEST and HIGH RANGE buttons, and rezero
the recording using the DC OFFSET switch and knob.
7) You
are now ready for recording. Set the amplifier METER to OFF and
withdraw the microelectrode and the ground pellet from the
bath. Carefully position the micromanipulator so that both you
and the microelectrode will be safe from damage. When you have
the crayfish in place and the muscles exposed (see next
section), you will reposition the microelectrode and ground
pellet in the bath, then run through steps 3-6 above to prepare
for recording. Each time you replace a microelectrode you will
again have to perform Steps 3-6.
E.
Crayfish Surgery
1)
Review the CD video guide for Crawdad Laboratory #4 - Crayfish
Muscle Resting Potential. Pay particular attention to section
4.2 which illustrates the proper technique for exposing the
superficial flexor muscles without damaging them. See also
Appendix A for explanatory figures and a background on the
crayfish superficial abdominal flexor system. A good video
of the dissection procedure is at:
http://www.wellesley.edu/Biology/Concepts/Html/crayfishabdomen.html
2)
Isolate and pin out a crayfish tail in a Sygard-lined dish, then
immerse it in cold crayfish Ringers. Choose a segment and
carefully expose the superficial flexor muscles on both
sides of the tail. Important things to watch out for in this
surgery are:
- use a fresh scalpel blade and keep it flat - DON'T
CUT TOO DEEPLY
- don't damage the muscle insertion line which runs
medio-laterally
- make sure that the IIIs (the third superficial
root) is intact and undamaged
If
any of these conditions are not met, then that half-segment
cannot be used.
3) Don't
proceed to the recording phase until you have at least three
intact half-segment muscle exposures. HAVE THE INSTRUCTOR CHECK
OUT YOUR SURGICAL RESULTS BEFORE PROCEEDING FURTHER.
II.
RECORDING AND MANIPULATING MUSCLE RESTING POTENTIALS
Once you have your electrode successfully tested and mounted,
and your crayfish suitably prepared, you will be ready to
penetrate and record from some muscle cells. Electrode
positioning is an acquired skill which takes some practice.
So practice. If and when you break off an electrode tip,
be prepared to start all over again with electrode preparation,
impedance determination, and capacitive compensation on your new
electrode.
As should be
apparent from a thoughtful examination of the Goldman expression
(see Crawdad Appendix B, Equation 8), the ion which exerts the
most influence over the resting potential is the one with the
highest resting permeability (and conductance), namely
potassium. It should also be fairly obvious that it is much
easier to manipulate extracellular (bath) concentrations than
intracellular concentrations. In the exercise sequence outlined
below, and detailed in Crawdad Lab #4, you will first measure
resting potentials in several muscle cells under standard
crayfish Ringers. You will then measure resting potentials under
baths containing increasingly high concentrations of potassium.
In several of the simulations in this course, you have modeled
the effects of directly injecting current into a cell, in order
to look at the membrane response. In the final part of
this section you will actually do just that. In fact, you
will be injecting current through the same electrode which you
are using to record the membrane potential. In order to do
this, you will need to have the amplifier compensate for the
current itself, as seen by the electrode. Fortunately, the
amplifier has as separate set of circuitry and controls for
doing this. |
A.
Baseline Resting Potentials
1) You
will be following the general methodology of Crawdad Lab #4. If
you have not already done so, review the CD videos associated
with this lab. Also read through the section of Appendix D
dealing with intracellular electrodes (pp. 52-53)
2)
Prepare a crayfish as described above. Mount a microelectrode.
Zero the Neuroprobe DC offset, check the electrode impedance,
and adjust the capacitive compensation. Start continuous
repetitive sweeps on the PowerLab.
3)
Position the microelectrode tip in close proximity to exposed
muscle fibers. You will not be able to precisely see the actual
electrode tip, so you must rely on the amplifier output as a
guide to fine positioning and cell penetration.
4)
Slowly advance the microelectrode tip using the fine control
knob on the micromanipulator. Watch the PowerLab screen and
listen to the audio monitor. As the electrode tip contacts the
muscle membrane the flat 0 trace and the audio tone may start to
fluctuate. At this point you can either carefully continue to
advance the tip until the cell in penetrated, or try to "buzz"
into the cell using the CAP. OVERRIDE button. When you enter
the cell the voltage on the PowerLab screen and on the
Neuroprobe LED meter should suddenly drop to a negative value,
the resting potential. This will be accompanied by an abrupt
rise in the tone of the audio signal. If the electrode is
making a tight seal with an undamaged cell the resting potential
should be stable and at least -50 mV in amplitude.
Q2:
Why is the resting potential negative? Are there any possible
bath conditions under which you might expect to record a
positive resting potential? Examine the Nernst and Goldman
equations if you are having trouble answering this.
Q3:
What would you expect to see in the way of a resting potential
if a cell is damaged and "leaky" or if the membrane does not
seal tightly around the electrode? |
|
|
Data Sheet Item
#2:
Print out a pair of superimposed
Scope traces showing measured electrode potentials "before" and
"after" penetrating a muscle cell. Indicate on the printout the
value of the resting potential for your cell. |
|
|
5)
Measure and record the resting potentials of at least five
muscle cells. The easiest way to reposition your electrode is
to back it out of the cell at least two full turns of the fine
advance control knob, turn the horizontal positioning knob a few
degrees, then carefully readvance the electrode until a new cell
is penetrated. Be sure to back the electrode out a safe
distance when you are through.
B.
Dependence of the Resting Potential on Extracellular Potassium
Concentration
1) Find
the stock solutions of normal (5.4 mM K+) and
high-potassium (60mM K+) Ringers. Prepare 200 ml
samples of each of two new intermediate-potassium solutions - a
2:1 mixture of normal:high-K+ Ringers and a 1:2
mixture of normal:high-K+ Ringers. Keep these
solutions on ice until they are needed.
Q4:
What are the potassium concentrations of these two new
solutions?
2)
Measure and record resting potentials for at least five muscle
cells under high-K+ Ringers and under each of the two
intermediate-K+ solutions. When you are finished,
record five more resting potentials under standard Ringers as a
baseline comparison to your original measures. The safest way
to change the bath solution is to remove all of the fluid with a
50 cc syringe, then slowly replace it with the new
solution. |
|
|
Data Sheet Item
#3:
Produce a table of your data
showing bath potassium concentrations (columns) and measured
resting potentials. At the bottom of each column enter the mean
resting potential under that condition and the standard
deviation.
|
|
|
|
|
|
Data Sheet
Item
#4:
Produce a plot of resting potential
as a function of the log external potassium concentration. You
may either plot all of your data points (scatterplot) or the
mean values with standard deviation or standard error bars. If
your relationship is a linear one, draw a best-fit regression
line. |
|
|
Q5:
Why should the resting potential vary as the log[K+]
o, rather than simply as [K+]o
itself?
Q6:
Should the membrane resting potential be more sensitive to [K+]
o, or [Na+]o? Why?
III.
PASSIVE MEMBRANE RESPONSE TO CURRENT INJECTION
In several of the simulations in this course, you have modeled
the effects of directly injecting current into a cell, in order
to look at the membrane response. In the final part of
this laboratory exercise you will actually do just that.
In fact, you will be injecting a square-wave current pulse
through the same electrode which you are using to record the
membrane potential. In order to do this, you will need to
have the amplifier compensate for the current itself, as seen by
the electrode. Fortunately, the amplifier has as separate
set of circuitry and controls for doing this. |
A Preparing the Amplifier for Current Injection
1)
Connect the CURRENT output of the Neuroprobe amplifier to to the
CH2 input of the PowerLab. Disconnect the CH2 input from the
Audio Monitor switch. Connect the + OUTPUT of the PowerLab to
the CURRENT GATE input of the Neuroprobe amplifier.
2) In
Scope, open the Display; Computed Functions . . . window and set
the display for Channel A and B. Turn on Channel B and set the
range to 200mV. Under the Setup; Sampling menu choose
Repetitive. Under the Setup;Stimulator set the stimulus to
produce a single pulse with a 5 msec delay, 10 msec duration,
and 4 volt amplitude. This pulse will trigger or “gate” an
actual current injection into the cell by the Neuroprobe
amplifier.
3) In
order to both inject current and measure a true cellular
membrane voltage response simultaneously through a single
electrode, you will need to “null” the response of the electrode
itself to the current pulse. The following steps will
accomplish this.
a) Make
sure that a good (.5-1.0 Mohm)
electrode and the reference ground pellet are both in the bath.
Set the DC balance to zero and adjust the capacitive
compensation to null out passive electrode properties, as in
section I.D. above.
b) Start
repetitive 50 msec sweeps on Scope.
c) Set
the Neuroprobe meter to CURRENT. Push in the HIGH RANGE
button. Turn the CURRENT knob to 0 (completely
counterclockwise). Turn the inner TRANSIENT knob slightly
clockwise (off of CAL) to activate current injection. Set the
CURRENT POLARITY switch to POS(itive).
d) Turn
the CURRENT knob clockwise to produce a 50 nA current pulse, as
monitored on the Neuroprobe meter. You should see square pulses
on both channels of the Scope display. Channel B displays the
current pulse itself at 1mV/nA. Channel A displays the
electrode response in mV.
e) Null
out the electrode response by first turning the DC BAL knob
reduce the center of the Channel A pulse trace down to zero.
Then adjust the inner and outer TRANSIENT knobs to minimize the
capacitive transients at both ends of the pulse.
f) Turn
the CURRENT knob back down to zero (counterclockwise).
B. Injecting Current
1)
Carefully advance the electrode into the cell and establish a
resting potential for that cell. Note – the resting potential
will show up on Scope Channel A. The Neuroprobe meter should
still reflect the injected current level, which should be zero.
2) Turn
the CURRENT knob up (clockwise) to inject a 10 nA current pulse,
as indicated on the Neuroprobe meter. Save one scope trace,
showing the cell membrane response. Now flip the POLARITY
switch to NEG(ative) and record a second, superimposed trace.
Produce similar traces for a positive and negative 20 nA pulse.
Turn the CURRENT knob back down to zero and the TRANSIENT knobs
back to CAL when you are finished.
3)
Examine, zoom, measure, and calculate to make the following
measurements for each trace:
10nA -10nA 20nA -20nA
a)
Amplitude I of the injected current pulse ____ ____ ____ ____
b)
Amplitude V the membrane response in mV ____
____ ____ ____
c)
Time constant t
of the rising phase (time to 63% max)
____ ____
____ ____
d)
Effective membrane resistance R (from V=IR) ____
____ ____ ____
e) Effective membrane conductance g (from g=1/R)
____ ____ ____
____
f) Effective membrane capacitance C (from
t
=RC) ____ ____
____ ____
Q7:
Is the membrane response symmetrical for positive (depolarizing)
and negative (hyperpolarizing) current pulses? If so, the
membrane response is said to be “nonrectifying”.
Q8:
Are your values for the membrane time constant, resistance,
conductance, and capacitance the same for different size pulses?
Q9:
Does the muscle cell behave like the simple parallel resistor
and capacitor circuit which we have been using to model it? |
|
|
Data Sheet
Item
#5:
Print out a single, well-labeled Scope trace showing the
superimposed responses to all four current pulses as described
in #5 above. Include the results of the calculations from #6.
|
|
|
III.
MINI-INDEPENDENT STUDY
Conduct a
small independent study of your own design involving the
dependence of resting potential or current injection responses
on environmental conditions. Crawdad Lab #4 provides some
ideas, such as bath temperature dependence, extracellular sodium
concentration, adding ouabain to the bath to block the Na/K
membrane pumps, or recoding from a single cell while carefully
changing bath ion concentrations. |
|
|
Data Sheet
Item
#6:
Produce well-annotated printouts which describe and illustrate the
results of your mini-independent study. |
|
|
IV.
SHUTTING DOWN
1) Make
sure that you have saved all of your data to the hard drive,
then quit Scope. Turn off the PowerLab box.
2) Turn
off the Neuroprobe amplifier.
3)
Properly discard all microelectrodes.
4) Flush
out the half-cell electrode holders with distilled water, then
air and store them dry. Flush out your microfil fiber needle
with distilled water, then air.
5) Make
sure that both the microscope and fiber-optic lights are turned
off.
6) Make
sure that both micromanipulators are magnetically secured to the
steel plate.
7) Make
sure that the Microelectrode R/C Meter is turned off.
8)
Return all solutions to the refrigerator and store all crayfish
parts in the freezer.
V. PREPARATION OF
THE LAB DATA SHEET
|
|
|
Your data sheet
should include at least FOUR of the items described in the boxes above.
Make sure
that
the axes of all of the graphs and print-outs are labeled and
calibrated. You should certainly discuss your results and the answers
to the questions with your partners and others in the lab. However,
please work independently when you prepare your data sheet.
|
|